Technical Information & parts

Hardware Eclipse black box for USB1.1 PC connection
Database OtoAccess™
Stimulus rate 40 or 90 Hz
Transducer Ear Tone ABR insert phone/TDH39 headphones
Level 0– 100 dB nHL in 5 dB steps.
Tone Burst Frequency 500, 1000, 2000, and 4000 Hz, both ears same time.
Bandwidth 1 octave ± ½ octave – 3 dB
Masking White noise 0 – 100 dB nHL
Analysis Time 6 minutes to detect a ASSR signal – can be extended up to 15 minutes
A/D Resolution 16 bit
Artefact Reject System Standard voltage based system
Gain 74 – 110 dB. Auto or Manual selection. (5µV to 3200µV input)
Channels 2, with separate detection algorithm
Accuracy method 1% or 5%
Gain Manual 5, 10, 20, 40, 80, 160, 320, 640 uV input
Anti aliasing filter Analog 8kHz, 24 dB / octave
Display Stimuli level and frequency
Session status
Expert or Basic View
Customer selectable correction factors available
Display Gain General Display Gain , applicable during testing. Display Gain, applicable during testing.
Controlled parameter Independent control of up to 8 simultaneous stimuli (max 4 per ear)
Independent start, stop control for each of the 8 stimuli
Stimulus level control for each of the 8 stimuli
False pass probability 1 or 5%
Test protocols included for children and adult
Data I/O USB 1.1
Printout Hardcopy or as pdf file for EMR
Customized printouts.
NOAH NOAH compatible (NOAH 3.6 or higher)
(Estimated Audiogram available for other NOAH modules)
Networks May connect to a network.
Subsequent sessions viewing from reader stations.
With optional software, even tests in progress may be monitored and controlled from any reader station in the network.
Med. CE Mark/FDA Yes/Pending

Included Parts

  • Eclipse
  • EPA4 Preamplifi er
  • ETB4 Standard electrode cable with buttons
  • ETU4 Universal electrode cable
  • ETR4 Electrode cable with re-usable electrodes
  • PEG15 Set of 25 single use pre-gelled electrodes
  • EarTone ABR Insert ear phones w/foam tips
  • 20 pcs. of Infant eartip (2 x 10)
  • SPG15 Tube of skin preparation gel
  • Electrode gel
  • Alcohol pads (100 pcs.)
  • USB cable 2m
  • Power cable
  • Software CD as ordered
  • Database Software (OtoAccess)
  • Operation Manual
  • CE Manual

Optional:

  • Medically approved Fujitsu Siemens PC
  • TDH39 Headset
  • UCO15 Optical USB cable for Eclipse (can be delivered with 1 or 5 meters
    USB extension cable).

Upgradable:

  • EP15 or EP25 ABR system
  • DPOAE20 Distortion Product OAE System
  • TEOAE25 Transient Evoked Product OAE System