| Hearing Aid Analyzer | EN 60118-0, EN 60118-7, ANSI S3.22. |
| | Stimulus Signal | Warble Tone, Pure Tone, Random noise, Pseudo random noise, Band limited white noise, Chirp, ICRA, Real Speech, any other sound file (automatic calibration available). |
| | Stimulation Intensity Range | 40-100 dB SPL in 1 dB steps |
| | Intensity accuracy | ± 1.5 % dB |
| | Stimulus Distortion | Less than 1 % THD |
| | | | Coupler Microphone Ranges | 40-145dB |
| | FF Loudspeaker Output | Max 6 W into 8ohms, max 10 W into 4 ohms. |
| | Telecoil drive in Test box | 10 – 100 mA/meter |
| | Battery Simulator | Standard types are selectable, Custom types within 1.1 – 1.6 V, 0 – 25 Ohm range. |
| | Frequency Range | 100-10000Hz. |
| | Frequency Resolution | 1/3, 1/6, 1/12 and 1/24 octave or 1024 point FFT. |
| | Test box | Built-in test box holds telecoil drive as well as special dual speaker set for checking directional microphone function. |
| | Test box | HIT module comes with a set of Test Protocols loaded. Additional Test Protocols can be designed by user, or easily imported into the system. |
| | Compatible Software | NOAH 3, OtoAccess™ |
|
|
Continue to Gallery  |